Abx Micros 60 User Manual
Download Horiba ABX Micros 60 - Technical Manual 2.en.es. Share & Embed 'Horiba ABX Micros 60 - Technical Manual 2.en.es' Please copy and paste this embed script to where you want to embed. ABX Micros 60 ABX Micros 60 is a fully automated hematology analyzer from ABX Diagnostics (renamed as Horiba ABX). Horiba ABX Micros 60 is a powerful tool best suited for use in any physician’s office lab, hospitals, and small clinics. One type of micros training manual is that which has a technical device. It may help explain 3700 POS MICROS USER MANUAL PROGRAMMING. Our nationwide network of abx micros 60 manual is dedicated to Micros Programming Manual The Micros POS system relies on a database that was programmed specifically for the previous restaurant. ABX Micros CRP 200 - Technical Manual - RAA028A 1–1 Section 01: Introduction 1. Working conditions 1.1. Environment The operation of the ABX Micros CRP 200 should be restricted to indoor location use only! Operation of the instrument at altitudes of over 2000 meters (6000 feet) is not recommended.
- Hematology
- Impedance technology, photometry
- Microsampling: Only 10 µL
- No compressor (stepper motor technology)
- Dedicated infectious disease sample containment for labs
ABX Micros 60 Concept And Technology
- Measurement methods: Impedance technology, photometry.
- Liquid valves (without pinching): Reliability and increased precision
- No compressor (stepper motor): Reliability, less maintenance and noise
Manufactured by HORIBA ABX SAS
Technical
Weight | 31 lbs |
Power supply | 90/240 V - 50/60 Hz - 230 VA. |
Dimensions | 16.5 in x 14 x 12.5 in (HxWxD). |
Printer | Laser Printer |
Throughout | 60 samples/hour (on open tube) |
Data management | RS 232 output. |
Memory | LiteDM Patient Data Management System |
Screen | 2 lines of 40 alphanumerical characters. |
Identification | Barcode reader (optional). |
Samples
60 samples/hour (open tube) | |
---|---|
Parameters | WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV, GRA%/#, LYM %/# and MON%/# |
Graphics | WBC, RBC and PLT histograms |
Sample-Volume | CBC mode: 10 µL |
Stability | 24 hours (CBC) |
Operating mode | Open tube mode with 4 position tube holder (Cal, 125 or 500 µL, 3 or 5 mL) |
Identification | Patient I.D., Sequence# |
Flags | Pathological flags (WBC, RBC, and PLT) normal limits and counting failure flags |
Results
Parameters | Precision (cv %) | Linearity |
---|---|---|
WBC | < 2,5% | 0-100 x 10³/mm3 |
RBC Cool edit pro 2.0 for macor mac. | < 2% | 0-8 x 106/mm3 |
HGB | < 1.5% | 0-26 g/dl |
HCT | < 2% | 0-80% |
PLT (whole blood) | < 5% | 0-2200 x 10³/mm3 |
PLT (PLT concentrate) | < 5% | 0-4000 x 10³/mm3 |
- Documents
- Videos
- News
DESCRIPTION
8 or 18 fully automated parametersCBC / 3 Part DIFFOpen or Closed TubeImpedance technology, photometry
Microsampling: Only 10 µL
Smart card based memory storage (optional)
No compressor (stepper motor technology)
FORUMSView All (20)
Ask a New Question2Replies5 months ago | 5 months agoHavingproblem with RDW, Low results Having problems with low RDW results.RBC 5,09-HGB 15,1-HCT 49,3 MCV 97-MCH 29,7 MCHC 30,66 PLT 230 RDW 2,2. Reply |
-Mkama zebedayo a year ago | a year agoservicing micro 60 and screen master i would like to know what can i do to my micro 60 that is giving out high PLT reading that is going up to 22222 sometimes giving DIL.also to my screen master that is giving out a wide range of deviation to a single sample or QC testing sample aslo the failure of QC,How to calibrate this one since it has failed more than fifth times?Reply |
2Replies2 years ago | 2 years agoSERVICE MANUAL Dear All,Am looking for a service manual for Horiba ABX Micros 60. May anyone who has it share by sending to: vgmakwinja@medcol.mwThanks in advance.Reply |
SERVICE COMPANIESView All Hematology Analyzer Companies
FEATURES
Measurement methods: Impedance technology, photometry. Liquid valves (without pinching): Reliability and increased precision No compressor (stepper motor): Reliability, less maintenance and noise
Additional Specifications
Micros 60 Manual
Technical
Weight
31 lbs
Power supply
90/240 V - 50/60 Hz - 230 VA.
Dimensions
16.5 in x 14 x 12.5 in (HxWxD).
Printer
Dot matrix.
Throughout
60 samples/hour (on open tube)
55 samples/hour (closed tube)
Data management
Automated smart cards used for QC and patient data storage (optional).
RS 232 output.
Mono-directional connection.
Memory
Last result (with curves) or 78 results (optional smart cards based memory)
Screen
2 lines of 40 alphanumerical characters.
Keyboard
Barcode reader (optional).
Samples
60 samples/hour (open tube)
Parameters (16+2)
CBC Mode (18 parameters): WBC, RBC, HGB, HCT, MCV,*RDW, MCH, MCHC, PLT, MPV, PCT, *PDW, % and # of LYM, MON and GRA
Graphics
LMG histogram, RBC, PLT histograms
Sample-Volume
CBC mode : 10 µL
Stability
24 hours (CBC)
Operating sample mode
Open tube mode with 4 position tube holder (QC, 125 or 500 µL, 3 or 5 mL)
Identification
Patient I.D., Sequence#
Flags
Pathological flags (WBC, RBC, and PLT) normal limits and counting failure flags
* RUO (Research Use Only)